The instructor guide to ACLSMED™ PACER, the temporary pacemaker trainer: a pulse generator and a bedside monitor on one tablet, driven by a live heart model. Every dial, key and page, the lessons, the scored cases, and TEACH mode, where the class watches the monitor while students hold the generator.
A self-paced trainer for temporary transvenous and epicardial pacing with an external pulse generator. The learner sets the generator while watching a bedside monitor, and the heart answers beat by beat.
Training only. Not a medical device. It does not connect to any pacemaker, generator or patient, and must not be used for any decision about a real patient. It gives no medical advice.
The generator and monitor are our own design. Their keys, menus, alarm names and numbers match no manufacturer's device. Real generators differ: your unit's policy and the operator's manual for your generator always come first.
No sign-in, no server, nothing collected. It works offline.

Free run, nothing scored. Change the patient and add faults from the generator's TRAINER page.

Stepped lessons on the real dials, with SHOW ME and check questions. All free.

20 scored cases in four tiers with a full debrief, and ten-strip drill rounds: name it, then fix it.

The monitor at the front of the room, the generator in a student's hand. Every dial turn is applied on the monitor.
| § | Chapter | Read it when |
|---|---|---|
| 01 | Devices, plans & privacy | Choosing hardware, Free or Pro |
| 02 | The generator | Every dial, key, screen and page |
| 03 | The monitor | Leads, 12-lead, markers, SENSE VIEW, MENU, sound |
| 04 | Practice | Free run, patients and faults, the tour and DEMO |
| 05 | Learn | The 8 lessons |
| 06 | Scenarios | Running a case, scoring, the debrief, all 20 cases |
| 07 | Strip drill | Ten strips a round |
| 08 | TEACH mode | Two devices for a class |
| 09 | Running a class | Checklist and a session plan |
| 10 | Reference | Troubleshooting, FAQ, ranges, glossary |
← → or space to move · Home / End · P prints · F fullscreen. Add #n to the address to open slide n. The PDF is the same deck, one slide per page.
Each picture is a capture of the app on an iPad or an iPhone, not a mockup. Numbered markers on a figure match the numbered list beside it.
Medical advice. The numbers in the lessons and cases are this trainer's teaching defaults. Follow your unit's policy.

The right panel shows the learner's own progress: cases passed by tier, lessons completed, drill rounds and the case up next. Across the top: GET PRO (free tier only), ? Help, ⚙ Settings and i the training notice.
Lessons first, then Practice, then Scenarios and the Strip drill. Before any case is tried, the up-next key starts the next lesson.
A tablet runs the whole trainer. A phone is the generator a student holds. Pro is one purchase inside the app.
| Device | What it runs | TEACH role |
|---|---|---|
| iPad | The whole trainer, in landscape: Practice, lessons, scenarios, the drill, TEACH | Monitor or generator |
| iPhone | Only the TEACH generator. It opens on TEACH's role screen and nothing else; the monitor, lessons, scenarios and drill run on an iPad | Generator |
| iPhone Duo | Open: the whole trainer, as an iPad runs it. Closed: a screen that asks you to open it | Open: either role |
| Android tablet | The whole trainer, in landscape; a foldable opened flat counts as a tablet | Monitor or generator |
| Android phone | Only the TEACH generator, joined to an Android tablet on the same Wi-Fi | Generator |
iPad and iPhone pair with each other over Wi-Fi or Bluetooth. Android devices pair with each other over the same Wi-Fi. An iOS device cannot pair with an Android device (§ 08.02).




| Free | Pro, yearly | Pro, lifetime | |
|---|---|---|---|
| Price | USD 0 | USD 39.99 a year, renews automatically | USD 79.99 once, never renews |
| Practice and TEACH Practice | Included | Included | Included |
| All 8 lessons | Included | Included | Included |
| Scenarios | 4 cases: A1, A2, B1, C1 | All 20 | All 20 |
| Strip drill | One round a calendar day | No limit | No limit |
There is no trial period. The free tier does not expire.
ACLSMED™ PACER Pro is bought inside this app. It is not part of ACLSMED™ SIM Pro or the Suite.
In TEACH the scenarios keep their Pro gate on the monitor device. A phone holding the generator needs nothing.

The paywall opens from three places:
RESTORE PURCHASES is on the paywall and in Settings, beside Manage subscription. Settings shows "Pro is active on this Apple Account" once it is.
Scores and best results are kept on the device and shown with or without Pro. Locking never hides anything already earned.

A fresh install opens on this notice. The learner taps I UNDERSTAND once. It says:
Read it again any time from i on Home or Settings > Training notice. The device itself is printed TRAINING ONLY, NOT A MEDICAL DEVICE.
The app collects no data. It runs on the device and needs no connection. Only buying or restoring Pro talks to the App Store or Google Play.
The link carries dial positions, key presses, generator screen frames and the session PIN between two devices in the same room. Nothing is stored, logged or sent anywhere else.
iOS asks for local network access the first time TEACH starts. Deny it and the app still works in full on one device; only TEACH cannot find a partner.
Scenario best scores, drill rounds and completed lessons. Settings > Your progress > Reset progress clears them after two confirmations. Pro is not affected.
Hand-held, three dials, two screens. Our own generic device. Everything the learner sets lives here.
12345678On an iPad the generator takes the left third and the monitor the right two thirds. On a phone the generator fills the screen (§ 02.13).
The pointer and the engraved value follow. Each step clicks (Dial clicks, § 03.09).
One step up or down.
The value keeps stepping while a finger stays on it.
| Dial | Range | Steps | Factory |
|---|---|---|---|
| RATE | 30 to 200 ppm | 1 ppm to 100 · 2 ppm to 170 · 5 ppm to 200 | 80 |
| A OUTPUT | 0.1 to 20 mA | 0.1 to 1 · 0.2 to 2 · 0.5 to 5 · 1 to 20 | 10 |
| V OUTPUT | 0.1 to 25 mA | 0.1 to 1 · 0.2 to 2 · 0.5 to 5 · 1 to 25 | 10 |
The step gets finer near the small numbers. Between 1 and 2 mA the output moves in 0.2 mA steps, so a threshold of 1.5 mA reads 1.6 on the dial. The lessons and cases allow for this.
The top screen shows the mode in large letters with its meaning under it (DDD: "Paces A and V, senses A and V, inhibits and triggers"), then RATE, A OUTPUT and V OUTPUT in big numerals, the two sensitivities in mV, the AV interval, and the battery. Banners appear for PAUSED, LOCKED and EMERGENCY.
PACED % says how often the generator paced instead of sensed. The generator cannot confirm capture. Capture is read on the monitor: the complex after each spike, the PACER tile's CAPTURED, the pulse.
| Flag | Meaning |
|---|---|
| AP | Atrial pace (above the line) |
| AS | Atrial sense that reset a timer |
| VP | Ventricular pace (below the line) |
| VS | Ventricular sense that reset a timer |
| AR · VR | A sense inside a refractory period: seen, not acted on |
| N | Noise |
| bar | Joins an atrial event to the ventricular pace that followed (the AV interval) |
The strip scrolls the last 6 seconds. It is the teaching heart of the device: read it before you touch a dial.

| Key | Press | What happens |
|---|---|---|
| PAUSE | hold | No output and no sensing while held, so the patient's own rhythm shows. Lets go by itself after 10 s. Use with care in a dependent patient. |
| EMERGENCY | one press | DOO 80, A 20 mA, V 25 mA, whatever was set, even if the generator was off. The key then reads EXIT EMERG. |
| EXIT EMERG | press or change | Any change you make settles the emergency; the values stay until you change them. A second press leaves it too. Nothing is restored. |
| LOCK | hold 2 s | Every dial and key is ignored, the screen says LOCKED. Hold again to unlock. |
| MENU | press | Turns the pages of the lower screen. |
| ON / OFF | hold 2 s | Switches the generator off or on. Off: no output, dark screens. |




| Mode | The app's own one-liner |
|---|---|
| AOO | Asynchronous atrial pacing. Rare; testing, or electrical noise with intact AV conduction. |
| AAI | Sinus bradycardia or sinus arrest with intact AV conduction. Keeps AV synchrony with one lead. |
| VOO | Asynchronous ventricular pacing. Noise, or when sensing cannot be trusted and the patient is dependent. |
| VVI | The workhorse. AV block, slow AF, any bradycardia when one ventricular lead is in. |
| DOO | Emergency asynchronous AV sequential pacing. The EMERGENCY key sets it. |
| DVI | AV sequential pacing without atrial sensing. Historic; competes with intrinsic P waves. |
| DDI | Dual-chamber pacing without atrial tracking. Atrial arrhythmias with AV block. |
| DDD | Full dual-chamber pacing with tracking. AV block with a healthy sinus node: keeps the atrial kick. |
A tapped mode takes effect at once. The line under the keys repeats the one-liner for the mode chosen.

A SENSITIVITY 0.4 to 10 mV (factory 0.5) and V SENSITIVITY 0.8 to 20 mV (factory 2.0), each with − + and ASYNC, which switches sensing off on that channel. A row the mode does not use says NOT IN and the mode.

AV INTERVAL · UPPER RATE (its note shows where the 2:1 point falls) · PVARP · A REFRACTORY · V REFRACTORY · V BLANKING after an A pace. Rows a mode does not use are dimmed.

RAP RATE 80 to 800 ppm (factory 300). Hold the RAP softkey (the third key on this page): atrial pacing at the RAP rate, no sensing, the ventricular channel off. Let go and the previous mode returns. Used to overdrive atrial flutter (Lesson 6, case B5).

BATTERY %, LEAD CHECK A and V (OK or OPEN), pulse width 1.0 ms fixed, and the device's three tones, Alarm tone, Beat beep and Spike click, each ON or OFF with a level set by the − + softkeys.



| Patient |
|---|
| Sinus rhythm 75 |
| Sinus bradycardia 45 |
| Complete heart block, junctional escape 40 |
| Complete heart block, ventricular escape 35 |
| Complete heart block, no escape |
| Slow atrial fibrillation 45 |
| Atrial flutter 2:1 |
| Mobitz II 2:1 |
| Asystole |
| Complete block with retrograde conduction |
| Fault | What it does |
|---|---|
| A / V lead off | No output reaches the lead, nothing is sensed |
| A / V loose connection | Output and sensing drop out at random for 0.5 to 3 s |
| A / V lead dislodged | Threshold above any output; sensing falls |
| A / V threshold rise | Exit block: the threshold rises above the set output |
| Low battery | 10%, falling 1% every 20 s until depleted |
| Myopotential noise | Muscle bursts on the V channel (oversensing) |
| Electrical noise | Mains interference on both channels and the monitor |
| Generator failure | No output on either channel |
A fault stays on until it is cleared.
| Alarm | What to check (the app's own advice) |
|---|---|
| LOW BATTERY | Change the battery soon. Have a second generator at the bedside. |
| BATTERY DEPLETED | No output. Press EMERGENCY on the spare generator and connect the leads. |
| A LEAD OFF | Check the atrial lead at the terminal block and at the patient. |
| V LEAD OFF | Check the ventricular lead at the terminal block and at the patient. |
| NOISE A | Interference on the atrial channel. Check the lead, the connections and nearby equipment. |
| NOISE V | Interference on the ventricular channel. The generator paces asynchronously until it clears. |
| PAUSED | No output while PAUSE is held. Release it. |
| LOCKED | Controls are locked. Hold LOCK for 2 seconds to unlock. |
| EMERGENCY | DOO at 80 with maximum outputs. Settle the settings once the patient is stable. |
| HIGH RATE | Sensed ventricular rate over 150 for 10 seconds. Look at the rhythm and the upper rate. |
| OUTPUT AT MAXIMUM | Capture may still be lost. Check the lead position and call. |
| OFF | The generator is off. Hold ON/OFF. |






In TEACH the phone (or a second iPad) shows only the generator: its screens, dials and keys, full screen and upright. It runs no engine of its own. It mirrors the monitor's frames ten times a second, and every dial turn and key press goes back to the monitor and is applied there.
What the class watches. The surface ECG, the generator's own markers lined up underneath, and the numbers.
1234567| Key | Tap | Hold or notes |
|---|---|---|
| LEAD II | Row 1 to the next of the twelve leads. The label shows the lead. | Hold opens the lead picker for row 1 and row 2 |
| 12-LEAD | All twelve leads, three by four, with a 10 s lead II rhythm strip | Tap again to go back |
| FREEZE | Holds the traces; the status line shows FROZEN | Tap again to run |
| MARKERS | Shows or hides the PACER MARKERS channel | Lit when on |
| PATIENT | Practice: opens the generator's TRAINER page | TEACH: tap for patients, hold for faults on the instructor panel |
| IDENTIFY | Replaces PATIENT in lessons, scenarios and the drill: tick what the strip shows, SUBMIT | § 06.04 |
| SENSE VIEW | Swaps the pleth row for what the generator sees | § 03.05 |
| HELP | The alarms showing with what to check, what the strip can show, the two rules | § 03.07 |
| MENU | Sound, shell colour, DEVICE TOUR, HOME or LEAVE | § 03.08 |
| HOME | Back to Home | In TEACH it asks before leaving |



The marker channel lines the generator's AP, AS, VP and VS flags up with the ECG above. A spike can be matched to what the generator thought it was doing.
FREEZE holds every trace so you can point at a spike and its marker together. Lesson 8 uses it to show AP then P at 40 ms, VP then QRS at 30 ms.

SENSE VIEW swaps the pleth row for the electrogram the generator sees on its lead, in mV, next to the sensitivity number. What rises above the line is sensed.
Sensing is on the device, so what the class sees on the surface ECG and what the generator senses can disagree. That gap is the teaching.
Raise the sensitivity number until the marker strip stops showing AS or VS. That is the sensing threshold. Then set half of it.
The heart's rate on the ECG, in bpm. The generator's set rate and PACER DETECT are in ppm.
A cuff reading at the start of every session, then every 3 minutes with a "next in" countdown. A case with an arterial line shows ART live. Loss of the atrial kick shows here.
The pleth follows each captured or intrinsic beat.
Mode and rate, A and V output, CAPTURED % and PACER DETECT, the spike rate the monitor detects in ppm. Capture is reported here, never on the generator.
Keep the two words apart with learners. A generator pacing at 80 ppm with loss of capture can sit beside a heart rate of 35 bpm.

The monitor's HELP: ALARMS NOW, each with what to check; WHAT THE STRIP CAN SHOW, the ten findings; THE TWO RULES for sensing and capture.

Help from ? on Home or Settings: where to start, every control, capture and sensing, the modes, dual-chamber timing, troubleshooting, scoring, a glossary, and what is saved.

MENU and Settings write the same values, so they always agree. Letting go of a slider plays a sample at the new level. While SILENT is on, the rows are dimmed but still live.

Generator: shell colour; Practice opens DDD 80 CAPTURING, VVI 80 CAPTURING or GENERATOR OFF. Demo mode (§ 04.04). Guides: console tour, Help, training notice. Pro, Your progress, About.
| Sound | What it is | Default |
|---|---|---|
| Silent | Mutes every sound in the app | Off |
| Alarm tone | With a generator or monitor alarm, until silenced | 70% |
| Beat beep | The monitor beep on each beat | 35% |
| Spike click | A soft click with every stimulus, so capture is heard as well as seen | 50% |
| Key clicks | When a key goes down and comes back up | 50% |
| Dial clicks | For each step of a dial | 50% |
| Screen taps | When you tap a button on the screen | 50% |
| Slider ticks | As a slider moves | 50% |
Each sound has an on/off switch and a level. The device's volume buttons set the overall loudness.
Free run. Nothing is scored. The best place to demonstrate on a projector and to let learners play.

Practice opens pacing DDD 80 at 10 mA over complete heart block with a junctional escape of 40, capturing, sound on. Change the start in Settings > Practice opens.
The first Practice on a device starts the console tour; after that it is under MENU > DEVICE TOUR or Settings.
Drag the V OUTPUT dial down below 1 mA. The spikes lose their QRS and the rate falls to the escape. Turn it back up and capture returns. Then try PAUSE, EMERGENCY and the MODE page.




The tour spotlights one part at a time on a live Practice generator while the patient keeps running. NEXT and BACK step through it; SKIP ends it. Start it from MENU > DEVICE TOUR, TRAINER > SHELL > DEVICE TOUR, or Settings > Guides.
| # | Step | # | Step |
|---|---|---|---|
| 1 | Welcome | 11 | The four softkeys |
| 2 | Lead terminals | 12 | PAUSE |
| 3 | The top screen | 13 | EMERGENCY |
| 4 | The marker strip | 14 | LOCK, MENU, ON/OFF |
| 5 | RATE | 15 | The monitor: two ECG leads |
| 6 | A OUTPUT and V OUTPUT | 16 | The marker channel |
| 7 | The menu screen: MODE | 17 | Pleth, ART, or what the generator sees |
| 8 | SENSE | 18 | Numbers |
| 9 | TIMING | 19 | LEAD, 12-LEAD, FREEZE, MARKERS |
| 10 | RAP · DEVICE | 20 | PATIENT, SENSE VIEW, HELP, MENU, HOME |

For a booth, a classroom screen as people arrive, or a walk-through. The generator runs itself for about three minutes, pressing its own keys: set-up, capture, thresholds, sensing, the modes, PMT, EMERGENCY and the 12-lead. Then it loops.
The screen stays awake while it runs.
Stepped lessons on the live generator and monitor. About six to eight minutes each. All free.



| # | Lesson | Objective (from the app) | Min |
|---|---|---|---|
| L1 | What the generator does | Name the parts, turn it on, read the top screen and the marker strip while it paces a patient in complete heart block. | 6 |
| L2 | Rate and capture (VVI) | Set the rate, confirm capture, find the ventricular capture threshold and set the output with a safety margin. | 7 |
| L3 | Sensitivity and sensing (VVI) | Watch a sensed beat inhibit the generator, find the sensing threshold, set half of it, and see asynchronous pacing. | 7 |
| L4 | The modes | Read the NBG letters and try every mode on one patient with sinus bradycardia and intact conduction. | 7 |
| L5 | Dual-chamber timing | AV interval, upper rate limit and PVARP in DDD; Wenckebach and 2:1 upper-rate behavior and PMT, and ending each. | 8 |
| L6 | Atrial pacing and overdrive | AAI for sinus bradycardia, the atrial threshold, and rapid atrial pacing to overdrive atrial flutter. | 7 |
| L7 | Troubleshooting | Failure to capture, undersensing, oversensing and failure to output, and what to check in which order. | 8 |
| L8 | The paced 12-lead | RV apical pacing, the paced P wave, fusion and pseudofusion, where the spikes fall; then the EMERGENCY key. | 6 |
Twenty scored cases in four tiers. The patient and the room change; the learner reads the monitor and works the generator.

| Tier | Name | Cases |
|---|---|---|
| A | Set-up and thresholds | 5 |
| B | Modes and dual-chamber timing | 5 |
| C | Troubleshooting | 6 |
| D | Clinical management | 4 |
Each row shows the learner's best score. Free cases are A1, A2, B1 and C1; the rest wear a lock until Pro. A locked case still shows the best score earned in it.

Events describe the patient and the room: symptoms, what the nurse says, which alarm sounds. The learner reads the monitor and the marker strip for the finding.

Leaving a case before it ends saves nothing; it can be started again from the list.

IDENTIFY: tick everything the strip shows, then SUBMIT. The ten findings are the drill's (§ 07.02).

| CHECK | Leads and terminals · Battery · Patient · Pulse · Blood pressure · Chest X-ray · Connections and cable |
| ACT | Swap the generator · Change the battery · Secure the leads and connections · Transcutaneous pads as backup · Ask for the lead to be repositioned · Give atropine |
| CALL | Call the provider · Call electrophysiology |
Each step has a window. Done in time: full points. Done late: half. A step judged on a setting pays only after the learner has touched that control, so a setting already right at the start pays nobody.
Things that must be true before the case ends, such as capture held to the end or the provider called. Every one must be met to pass.
Three questions at the end, worth 4, 3 and 3. The answers are shown in the debrief.
Harmful actions, such as leaving a dependent patient without capture too long. Any one fails the case and caps the score at 59.
Score 80 or more, and every must-happen item met, and no never-event. A learner who does nothing scores 0. Only the best score of each case is kept.




The questions and the debrief appear on the monitor, so the whole class debriefs from the same screen (§ 08.07).
| ID | Case | Patient and setting | What the case teaches | Time | |
|---|---|---|---|---|---|
| A1 | Start pacing | 68 M · inferior infarct day 1 · new V wire, generator off | Turn on, VVI, rate 80, confirm capture, check pulse or BP | 4:00 | Free |
| A2 | Find the ventricular threshold | 74 F · complete heart block · VVI 80 at 10 mA | Lower to loss of capture, find the threshold, set a margin | 4:00 | Free |
| A3 | Find the sensing threshold | 61 M · after valve surgery · sinus rhythm, backup V wire | Raise the number until undersensing, find the threshold, set half | 4:00 | Pro |
| A4 | Dual-chamber set-up | 66 F · first evening after aortic valve replacement · both wires | DDD 90, both thresholds, both margins | 6:00 | Pro |
| A5 | Atrial threshold in AAI | 72 M · day 2 after bypass · atrial wire only | Atrial threshold and margin in AAI | 4:00 | Pro |
| B1 | Choose the mode | 58 F · day 1 after mitral repair · sinus bradycardia, set VVI 60 overnight | A mode that keeps the atrial kick, atrial capture, check BP | 4:00 | Free |
| B2 | Flutter during DDD | 70 M · day 3 after bypass · heart block, DDD 90 | Recognize tracking at the upper rate, change to a non-tracking mode | 4:00 | Pro |
| B3 | Pacemaker-mediated tachycardia | 64 F · transvenous A and V leads · DDD 70, ectopics | Recognize PMT and end it with PVARP | 4:00 | Pro |
| B4 | Upper-rate behavior during exertion | 54 M · day 2 after bypass · DDD 70, upper rate 100 | Recognize Wenckebach then 2:1, set the upper rate | 5:00 | Pro |
| B5 | Overdrive a flutter with rapid atrial pacing | 69 M · day 2 after bypass · new atrial flutter | Set the RAP rate, deliver the burst, back to a pacing plan | 5:00 | Pro |
| ID | Case | Patient and setting | What the case teaches | Time | |
|---|---|---|---|---|---|
| C1 | Loss of capture | 77 M · coronary care day 3 · VVI 80 at 3 mA | Recognize failure to capture, output up, check leads, call | 4:00 | Free |
| C2 | Undersensing | 63 F · day 2 after valve surgery · sinus rhythm returned | Recognize undersensing, set the sensitivity | 4:00 | Pro |
| C3 | Oversensing | 81 M · heart block with no escape · VVI 70 | Recognize oversensing, fix the sensitivity | 4:00 | Pro |
| C4 | Failure to output | 70 F · VVI 80 · back from the commode | Check the leads first, then secure them | 4:00 | Pro |
| C5 | Low battery | 79 M · no escape · generator running five days | Check the battery, swap or change it in time | 4:00 | Pro |
| C6 | Lead dislodgement | 66 F · wire placed yesterday · coughing | Recognize failure to capture, pads as backup, call for repositioning | 5:00 | Pro |
| D1 | Inferior infarct with complete heart block | 71 F · emergency department · hypotensive, wire just placed | Turn on, VVI 70 to 80, capture, check BP, call | 4:00 | Pro |
| D2 | Post-TAVR block | 84 M · six hours after transcatheter valve · both wires | Complete block arrives: rate, DDD, capture, check BP, call | 5:00 | Pro |
| D3 | Slow atrial fibrillation after bypass | 75 M · day 1 after bypass · slow AF, DDD | A non-tracking ventricular mode, rate, capture | 4:00 | Pro |
| D4 | Asystole with a wire in | 73 F · backup VVI 40 at 2 mA · both wires | EMERGENCY at once, then settle to a pacing plan with capture | 4:00 | Pro |
What is the pacer doing? Ten random strips a round. Name it, then fix it with the real controls.

Free: one round a calendar day. Pro: no limit. Leaving a round saves nothing.


| Finding | On the strip (the app's own words) |
|---|---|
| Normal capture and sensing | Every spike is followed by its complex, and every intrinsic beat resets the timing. |
| Failure to capture | Spikes with no complex after them. |
| Undersensing | Spikes land where they should not: the generator did not see the intrinsic beats. |
| Oversensing | Pauses with no spike: the generator saw something that was not a beat. |
| Failure to output | No spikes at all when there should be. |
| Fusion / pseudofusion | A spike on a QRS that was already starting. |
| Upper-rate Wenckebach | The AV interval stretches beat by beat until a P wave is dropped. |
| Upper-rate 2:1 block | Every other P wave is not tracked. |
| Pacemaker-mediated tachycardia | Pacing at the upper rate, each paced beat conducted back to the atrium and tracked again. |
| Competitive pacing | An asynchronous mode paces into a chamber with its own rhythm. |
Two devices, one class. One tablet is the monitor the class watches. A phone or a second tablet is the generator the students hold.
Runs the one heart and generator model, the scoring, and the instructor panel in its left third. The bedside monitor keeps the right two thirds, exactly as on one tablet.
The generator's screens, dials and keys, mirrored from the monitor ten times a second. No engine runs on it.
Every dial turn and key press on the generator is applied on the monitor through the same path a local key uses. Latency on a local network reads as immediate.
A scored case in TEACH scores the students' actions exactly as it scores one learner on one iPad, and the two devices can never disagree about the patient.
| Monitor | Generator | Link | |
|---|---|---|---|
| iPad (or open iPhone Duo) | iPhone, or a second iPad | Wi-Fi or Bluetooth, no network set-up | Pairs |
| Android tablet | Android phone, or a second Android tablet | The same Wi-Fi network | Pairs |
| iPad | Android phone | None | No |
| Android tablet | iPhone | None | No |
Both devices need the same version of ACLSMED™ PACER. If they differ the join is refused: "Update both, then join again."
Allow local network access on both devices the first time it is asked.
A second device is told "This monitor already has a generator".




After about 10 seconds the phone says: check both devices are on the same Wi-Fi and that ACLSMED PACER is allowed Local Network in Settings > Privacy & Security.
123456The monitor's own PATIENT key works too: tap for the patients, hold for the faults.




| You see | What to do |
|---|---|
| GENERATOR OFFLINE | Nothing yet. The generator rejoins by itself with the same PIN when it comes back. Nothing on the monitor is lost. |
| RECONNECTING | On the phone: its controls are greyed until the link returns. |
| RELEASE GENERATOR | Frees the seat at once. A linked device is sent back to JOIN; an away one just loses the seat. A seat with no connection frees itself after 30 s. |
| Already has a generator | Another device holds the seat. Release it, or wait 30 s if it is offline. |
| Wrong PIN | Read the four digits on the monitor again. |
| 5 wrong PINs | New joins are paused for 30 s and the panel says so. The device in the seat can still rejoin. |
| LINK FAILED | Check local network access in Settings, leave TEACH and start it again. |

LEAVE TEACH on the panel, or HOME on the monitor, asks "Leave teach mode? The generator device will be disconnected." During a case it adds "The case will end unscored."
Hold LEAVE for one second. A short tap only shows "HOLD LEAVE 1 S", so a student cannot leave by accident.
A phone that locks or switches apps briefly says hello again on return and rejoins without the PIN.
A checklist for the hour before, and one way to use the hour itself.
| Min | Activity | App |
|---|---|---|
| Arrival | Let the trainer run itself on the class screen | DEMO mode |
| 0–10 | Parts of the generator, the marker strip, capture | Lesson 1 or 2 on the class screen, a learner on the dials |
| 10–25 | Pass the generator. Each student turns one thing: rate, output to loss of capture and back, sensitivity, a mode | TEACH, Practice |
| 25–35 | Inject faults from the panel; the class finds them from the strip | TEACH, FAULTS tab |
| 35–50 | One scored case with the class on the generator, then the debrief on the big screen | TEACH, SCENARIO (C1 is free) |
| 50–60 | Questions; set the lessons and the drill for self-study | Learn, Strip drill |
Lessons run about six to eight minutes and cases four to six. Build your own plan around your unit's objectives and policy.
Before a student touches a dial, ask what the marker strip will show next. Then turn it and look.
Freeze the traces to point at a spike, its marker and the complex after it at the same moment.
Show the electrogram against the sensitivity line when the class argues about undersensing and oversensing.
Faults show their effect, not their name, and the expected step stays hidden until you PEEK. Let the class find it.
Spike click and Beat beep together let the room hear a stimulus with no beat after it.
To hand the generator role to another device, press RELEASE GENERATOR. The patient keeps running.
Troubleshooting, frequent questions, the generator's ranges, and a glossary.
| Symptom | Likely cause | Fix |
|---|---|---|
| Dials and keys do nothing | LOCKED | Hold LOCK for 2 s |
| Dark generator, no pacing | Generator off, or Practice opens GENERATOR OFF | Hold ON / OFF for 2 s |
| No sound | SILENT on, a sound switched off, or device volume low | MENU or Settings > Sound; the volume buttons |
| No TRAINER page | A lesson, case or drill is running | TRAINER is in Practice only; in TEACH use the panel |
| An iPhone shows only TEACH | By design: a phone is the generator | Run the trainer on an iPad or open iPhone Duo |
| No monitor found | Different networks, or local network access denied | Same Wi-Fi; allow it in Settings > Privacy & Security |
| iPad cannot see an Android phone | By design: iOS pairs with iOS, Android with Android | Use two devices of the same platform |
| Join refused: versions | The two apps differ | Update both, then join again |
| A case is locked | Not in the free set | Pro, or Restore purchases on a new device |
| STRIP DRILL: free round used today | Free tier allows one a day | Tomorrow, or Pro |
| "Saved results could not be read" | The results file on the device could not be read | Progress starts again from there; the old file is kept on the device and nothing else changes |
No. It is a generic generator of our own design. Keys, menus, alarm names and numbers match no manufacturer's device. Teach your own unit's generator from its operator's manual.
No. Everything runs on the device. TEACH uses the local network or Bluetooth between two devices in the room.
No account exists. A phone used as the generator needs nothing. Pro matters on the device that runs the cases.
No. One generator per monitor. Use RELEASE GENERATOR to hand the role to the next device.
Not in this version. TEACH runs Practice and Scenarios. Lessons and the drill are single-device.
Best scores, drill rounds and lessons are saved on the device that ran them, only there. Nothing is uploaded.
No. PACER Pro is a separate purchase inside this app.
Never. It does not connect to any pacemaker, generator or patient. Training only.
| Setting | Range | Steps | Factory |
|---|---|---|---|
| MODE | AOO · AAI · VOO · VVI · DOO · DVI · DDI · DDD | — | DDD |
| RATE | 30 to 200 ppm | 1 · 2 above 100 · 5 above 170 | 80 |
| A OUTPUT | 0.1 to 20 mA | 0.1 · 0.2 · 0.5 · 1 | 10 |
| V OUTPUT | 0.1 to 25 mA | 0.1 · 0.2 · 0.5 · 1 | 10 |
| A SENSITIVITY | 0.4 to 10 mV, or ASYNC | 0.1 · 0.2 · 0.5 · 1 | 0.5 |
| V SENSITIVITY | 0.8 to 20 mV, or ASYNC | 0.2 · 0.5 · 1 | 2.0 |
| AV INTERVAL | 20 to 300 ms | 10 | 150 |
| UPPER RATE | 80 to 230 ppm | 5 | 110 |
| PVARP | 150 to 500 ms | 10 | 300 |
| A · V REFRACTORY | 150 to 500 ms | 10 | 250 |
| V BLANKING | 20 to 60 ms after an A pace | 5 | 30 |
| RAP RATE | 80 to 800 ppm | 10 · 20 above 200 · 50 above 400 | 300 |
| Pulse width | 1.0 ms, fixed | — | — |
| ppm | Pulses per minute: the generator's set rate, and PACER DETECT. |
| bpm | The heart's rate on the ECG. |
| Pacing spike | The stimulus as the monitor draws it, a thin vertical line on every lead. |
| Capture | A stimulus that depolarizes its chamber. Loss of capture: a spike with no complex after it. |
| Capture threshold | The smallest output that captures every beat, in mA. |
| Sensing | The generator detecting the heart's own depolarization on its lead. |
| Sensitivity | The number in mV a signal must pass to be sensed. Smaller is more sensitive. |
| Inhibit · Trigger | A sensed beat stops the pace that was due. A sensed atrial beat starts the AV interval and a ventricular pace follows. |
| Asynchronous | Pacing at the set rate with no sensing (AOO, VOO, DOO, or ASYNC on a channel). |
| AV interval | Atrial event to ventricular pace, in ms. |
| PVARP | The atrial refractory period after a ventricular event, in ms. |
| Upper rate limit | The fastest tracked ventricular paced rate in DDD. |
| Fusion · Pseudofusion | A spike landing on a beat already starting; a spike on a complex it did nothing to. |
| Exit block | The threshold at the lead tip has risen above the output. |
ACLSMED™ PACER is education software for a simulated patient. It is not a medical device, it does not connect to any pacemaker, generator or patient, and it must not be used for any decision about a real patient. The generator is a generic teaching generator; real generators differ, and your unit's policy and the operator's manual for your generator always come first. It does not give medical advice. In a real emergency, call your local emergency number.